Radiation Therapy
Cylindrical Ion Chambers
Victoreen Model 550 Series
•
Cylindrical Ion Chambers for use
with Model 35040 and Model 530
electrometers
Introduction
Model 550 Series Cylindrical Ionization
Chambers are fully guarded thimble-type
probes for use in a variety of applications.
Equilibrium caps are available for each
probe, extending their energy range up to
12 MeV. Sensitive volumes have been
selected such that the nominal coulomb-
to-roentgen calibration factors are
convenient powers of ten. The cables may
be terminated by triax BNC connectors for
compatibility with Model 530, Model
35040 electrometers, or coax UHF
connectors for compatibility with the older
model electrometers.
Applications
The family of Cylindrical Ionization
Chambers is designed to meet a wide range
of radiation measurement and dosimetry
applications required for diagnostic x-ray
and radiation therapy in medical facilities.
The 550 Series Ion Chambers offers volumes of 330, 33, 3.3, 0.33 cm
3
for scatter, dose
and rate measurements and accommodate energies from 21 keV to 12 MeV when used
with equilibrium caps.
The selection guide on the next page will assist the user in selecting the appropriate
chamber for rates from 0.03 to 2300 R/s and for sensitivities from 100 to 0.1 nA/R/s.
•
Wide range of applications in
Diagnostic X-Ray and Radiation
Oncology suites
•
Accommodates energies from
21 keV to 12 MeV (with
equilibrium cap)
•
The 550 Series offers volumes of
330, 33, 3.3, and 0.33 cm
3
•
Chamber sensitivities are
available for 100, 10, 1, and
0.1 nA/R/s
•
These chambers accommodate
rates of 0.03, 1.5, 10, or 2300 R/s
•
Ion Collection Efficiency
information is available for specific
frequencies
Specifications
Available model(s)
Equilibrium caps - model numbers and
specific energies
550-3 550-4 550-5 550-6A
550-3
550-4
550-5
550-6A
Scatter, Diagnostic
Orthovoltage, Supervoltage
Teletherapy, Orthovoltage
Supervoltage, Diagnostic
Applications
137
Cs,
60
Co,
2 MeV
550-3-26 550-4-26 550-5-26 550-6-25
4 MeV 550-3-27 550-4-27 550-5-27 550-5-26*
6 MeV 550-3-28 550-4-28 550-5-28 550-5-27*
8 MeV 550-3-29 550-4-29 550-5-29 550-5-28*
10 MeV 550-3-30 550-4-30 550-5-30 550-5-29*
12 MeV 550-3-31 550-4-31 550-5-31 550-5-30*
* Used in conjunction with 550-6-25.
Volume
330 cm
3
33 cm
3
3.3 cm
3
0.33 cm
3
± 6%, 21 keV to 520 keV
without build-up medium
± 3%, 42 keV to 520 keV
without build-up medium
Energy Response
Sensitivity
100 nA/R/s
10 nA/R/s
1 nA/R/s
0.1 nA/R/s
**Intensity Limit @ 99.5% (Rate)
0.03 R/s
1.5 R/s
10 R/s
2300 R/s
Ion Collection Eff. (Pulse)
0.3 mR @ 60Hz
2 mR @ 400 Hz
6 mR @ 1.3 kHz
80 mR @ 17kHz
Wall Material
Acrylic
Acrylic
Polystyrene
Polystyrene
Wall Thickness (density thickness)
166 mg/cm
2
166 mg/cm
2
133 mg/cm
2
133 mg/cm
2
Chamber Diameter
55.9 mm
28.2 mm
15.5 mm
7 mm
Cable Length
3 m
3 m
3 m
7.5 m
Max. Rep. Rate
60 Hz
400 Hz
1.25 kHz
17 kHz
*Stem Scatter-X-Ray
0.10%
0.70%
0.90%
0.70%
*Stem Scatter -
137
Cs
0.10%
0.20%
0.20%
0.20%
Distance for 1.5% Geometry Error
36 cm
18 cm
6 cm
6 cm
Leakage
4 x 10
-15
A
4 x 10
-15
A
4 x 10
-15
A
4 x 10
-15
A
Cable Termination: Triax BNC or Coax UHF. Specify when ordering.
* Stem Scatter is included in calibration.
** Rates are calculated from chamber geometry. For measurements of rates equal to or greater than those in the
above table, the chamber ion collection efficiency should be investigated.